Structural flaws are hiding in plain sight, and a shift is long overdue.
By Pujitha Gourabathini, BD

Pujitha Gourabathini is a quality and patient‑safety leader at Becton, Dickinson and Co. [Photo courtesy of BD]
ISO 14971 and its companion guidance document, AAMI TIR24971, both acknowledge that estimating benefits carries uncertainty, but so does the initial risk estimation during design and development. Yet the two are not treated equally.
Risks are evaluated quantitatively using severity scales, probability of occurrence numbers and risk level classifications. Benefits, on the other hand, rely on qualitative justifications and subject matter expertise.
In practice, this means manufacturers are comparing and justifying risk matrix numbers against a narrative of clinical value. The result is a benefit-risk analysis that is interpretive at best and, at worst, a documentation exercise that satisfies auditors without truly informing design decisions.
But since benefit has no formal quantitative structure, manufacturers default to conservative interpretations of acceptability because there is no structured way to argue otherwise. As a result, devices with meaningful clinical value can be late to market, over-restricted, or unnecessarily modified because the benefit side of the equation was never given the analytical weight it deserves.
What needs to change
The medical device industry needs to apply the same analytical discipline to benefits as it does for risks. Severity and probability scales are evaluated from system test data during design development and through field monitoring or manufacturing data after production release. The same logic applies to benefits.
Clinical evaluations, premarket studies, postmarket surveillance data, and equivalent device literature are all sources from which benefits can be evaluated quantitatively. Data collection is already part of the postmarket surveillance process, but manufacturers must take one step further to incorporate it into a structured benefit evaluation that is equivalent to a hazard analysis format.
By doing so, benefit-risk determinations don’t continue to remain narrative-driven estimates while there is substantial evidence available for a rigorous assessment. This thought process requires not just a new set of risk management tools, but a change in how risk management teams think about risk acceptability from benefit-risk perspective.
A proposed operational framework
A quantified benefit framework should mirror the architecture of a risk assessment.
Begin with a strength scale. This is essentially a ranking scale built on dimensions like breadth of patient impact, longevity of the clinical outcome, degree of health or quality-of-life improvement, procedural ease compared to alternatives, and positioning against the state of the art. These criteria are not meant to be abstract, but to be mapped directly onto the kinds of endpoints already captured in clinical evaluations and usability studies.
Then, layer on a probability scale, which is the likelihood that the device truly delivers the identified benefits across the intended population, drawn from clinical data that is available at that particular stage of the product lifecycle. Prior to market release, this may be inferred from literature on equivalent devices or early feasibility analysis data. Postmarket, these values will be evaluated from real-world evidence such as complaint data, registry outcomes, and postmarket clinical follow-up (PMCF) studies.
From the above two scales, derive a benefit level, which mimics a risk matrix or is formatted as a classification right next to the risk level in the acceptance matrix. By defining acceptability thresholds explicitly, the manufacturer can clearly assess what benefit level is required to offset a given risk level, and under what evidentiary context. This ensures that the “acceptability” arguments made from a benefit-risk analysis are backed by traceable and auditable determinations.
More importantly, this framework must be treated as a living document. Just as a hazard analysis is revisited periodically and updated with postmarket data, the quantitative benefit analysis should also be reviewed on a periodic basis. This practice will ensure compliance with post-market surveillance and PMCF obligations under EU MDR, turning a regulatory requirement into a substantial input to ongoing risk management. Benefit re-evaluation at defined intervals also creates a feedback loop that strengthens the safety claims and arguments from the risk management file over the total product lifecycle, and not just at the point of initial market entry.
Final thoughts
The goal of medical device risk management is not only to assess risks, but also to ensure that the benefit outweighs the risk. ISO 14971 provides fundamental architecture for that determination, but leaves a level of uncertainty in implementation.
A quantified, data-driven benefit framework closes that gap by replacing literature with structured evidence and making “acceptable” a term that can be well defended. In a regulatory environment that is only growing more demanding, and where postmarket accountability is expanding in all geographic regions, this shift is long overdue.
Pujitha Gourabathini is a quality and patient‑safety leader at Becton, Dickinson and Co. (BD), where she directs risk‑management strategy for the Medication Management Solutions business. She leads initiatives that strengthen lifecycle safety, elevate engineering‑for‑safety practices, and build regulatory confidence through robust application of ISO 14971, safety assurance cases, and evidence‑based risk frameworks.
Read more MDO Contributions and learn how to submit your own
The opinions expressed in this blog post are the author’s only and do not necessarily reflect those of Medical Design & Outsourcing or its employees.